The genetic probe reagent box may present the wrong result
Recall the company: Genetic probe (Gen-Probe) Company
Recall the products: 1)AccuProbe Group B streptococcus trained, appraised reagent box, 2820
2)The branch bacillus of tuberculosis compounds and trains appraising reagent box, 2860
3)Bird's tuberculosis branch pole compounds and trains appraising reagent box, 2835
Recall the range: 1)The lot number is 554077, 551345, 556007, 556185, 555368, 554864, 556421, 555167, 555096, 555166, 556184 and 554524, and the international lot number is the products of 554076, 555622.
2)The lot number is 554628, 555170, 556204, and the foreign lot number is the products of 554840.
3)The lot number is the products of 555767.
Recall the rank: Grade
Recall the reason: Reagent quantity of test tube may accord with the standard, may lack too in the reagent box influenced, some of test tube reagent quantity also may be empty, this negative result that may make the mistake.
Recall the measure: The genetic probe company sends out and recalls the notice to the customer, propose stopping using the products of the above-mentioned batches. Have already used product of this lot number and obtained the customer of the negative result as to those, advise them to notify the doctor negative results may be wrong. The genetic probe company advises them to isolate the products of the relevant influenced lot number, consult and destroy all surplus products in stock, and be sure to mark out the quantity of destroying products on customer's receipt table.
Sponsor: Newspaper office of the health
Do jointly: National food medicines supervise administration bureau medicines appraising the centre
National medicines bad reaction monitoring center
|